DNA-AI-222 |
|
12-Channel, fully isolated RTD/resistance input board Producto descontinuado |
|
|
|
The DNA/DNR-AI-222 are 12-channel fully isolated, simultaneously sampling RTD input boards compatible with Cube and RACKtangle chassis respectively. The DNA/DNR versions are electronically identical. The DNx-AI-222 features a 100 µA excitation current and 24-bit A/D converters. At the gain of 16 used 100 Ohm RTDs the AI-222 provides an input resolution of better than 0.005 °C or 0.002 Ohm in resistance measurement mode.
An A/D per channel configuration allows simultaneous sampling at rates up to 100 S/s each (1200 S/s aggregate). The A/D per channel configuration virtually eliminates input cross talk and channel settling time issues.
The DNx-AI-222 is an ideal, high accuracy RTD measurement device, offering total errors of less than 0.2 °C on all standard 100 Ohm RTDs. This accuracy combined with the 350 VDC/Vrms channel-to-channel and channel-to-chassis isolation makes the board a perfect solution for industrial temperature measurement, even when using non-isolated RTDs in high voltage environments.
The software included will perform all required RTD linearization and compensation and returns data in °C, °K, °F or °R when desired. The software also allows the temperature conversion to be disabled and the inputs can measure resistance in Ohms up to 40 kOhms.
Software included with the DNx-AI-222 provides a comprehensive yet easy to use API that supports all popular Windows programming languages as well as supporting programmers using Linux and most real-time operating systems including QNX, RTX, InTime, VXworks and more. The UEIDAQ Framework supplies complete support for those creating applications in data acquisition software packages such as LabVIEW, MATLAB/Simulink, DASYLab or any application which supports ActiveX or OPC servers.
|
|
Characteristics: |
- For use in “CUBE” chassis
- 12 fully differential analog input channels
- 2, 3 and 4 wire connections
- 100 Ohm DIN and US measurements
- Channel-to-channel and channel-to-chassis isolation
- 0.005 °C resolution, 0.2 °C accuracy
- 0 to 40,000 Ω resistance measurement range
- Built-in 50, 60 and 400 Hz rejection (at select sample rates)
- 100 Hz, per channel sample rates
|
|
Technical Specifications |
Parameter |
Number of channels: |
12 fully differential plus |
ADC resolution |
24 bits, < 0.005 °C with 100 Ω PT RTDs
< 0.002 Ω in resistance measurement mode |
Sampling rate |
up to 100 samples/sec per channel
1200 S/S board aggregate |
Measurement error
0 to 4999 Ω
5000 to 40,000 Ω |
0.2 °C Max (100 Ω PT RTDs)
0.1 Ω Max (resistance mode)
tbd Ω Max (resistance mode) |
RTD excitation current |
100 µA typical (open TC source off) |
Resistance range |
0 - 40,000 Ω |
General A/D specifications |
Input offset |
< 3 µV @ 25°C, (<15 µV -40°C to +85°C) |
Gain error |
±0.005 % (typical) |
Input INL error |
6 ppm typical, 15 ppm max |
Input impedance |
>5000 MΩ |
Input range |
±2.048 Volt (gain = 1) |
Gains |
1, 2, 4, 8, 16, 32, 64 |
Anti-Aliasing filtering |
@47.6% of sample rate, ~100 dB/decade |
50/60/400 Hz notch filtering |
>70 dB at sample rate = 20 Hz or less |
Chan to Chan crosstalk |
< 0.5 µVrms |
Isolation |
350 Vrms, chan-to-chan & chan-to-chassis |
Overvoltage protection |
-15 V to +15 V (power on or off, current must
be limited to ±20 mA) |
Power consumption |
4 W max |
Operating temp. (tested) |
-40°C to +85°C |
Operating humidity |
95%, non-condencing |
Vibration |
IEC 60068-2-6 |
5 g, 10-500 Hz, sinusoidal |
IEC 60068-2-64 |
5 g (rms), 10-500Hz, broadband random |
Shock |
IEC 60068-2-27 |
50 g, 3 ms half sine, 18 shocks @ 6 orientations
30 g, 11 ms half sine, 18 shocks @ 6 orientations |
|