DNR-AI-207 |
|
16-channel, 18-bit, 1 kS/s per channel, analog input, data acquisition board for the PowerDNR RACKtangle® |
|
|
|
The DNR-AI-207 is a 16-channel A/D board featuring 16 differential input channels, 18-bit resolution, and 12 software selectable input ranges. Each channel may be sampled at up to 16 kS/s, though the maximum board throughput is 16 kS/s so applications monitoring multiple channels will reduce the per channel sample rate (e.g. 1 kS/S for 16 channels, 2 kS/S for 8 channels, etc.). Additionally, the DNx-AI-207 provides a dedicated CJC channel that can be used for reading from the built-in CJC sensor on the DNA-STP-AI-U or AI-207TC terminal panels. When used with DNA-STP-AI-U or 207TC panel, the DNR-AI-207 layer also features a direct connection to RTDs or thermocouples (with open TC detection). The software included will perform all required RTD/TC linearization and CJC compensation and return data in °C or °F if desired.
Another key feature of the DNx-AI-207 is the oversampling engine, allowing the board to automatically acquire as many samples as possible for the given gain/speed and average them, thus dramatically improving the noise floor. |
|
Characteristics: |
- 16 differential channels
- Additional dedicated CJC channel
- Maximum sampling rate of 1kHz per channel
- 18-bit resolution
- ±10V input range
- Gains - 1, 2, 4, 8, 10, 20, 40, 80, 100, 200, 400, 800
- Overvoltage protection (-40V to +55V)
- Direct Inputs for thermocouples (with DNA-STP-AI-U panel)
- Embedded averaging engine
|
|
Technical Specifications |
Parameter |
Number of channels: |
16 fully differential plus
1 single-ended dedicated CJC channe |
Programmable DIO line |
1 (external trigger) |
ADC resolution |
18 bits |
Sampling rate |
1 S/s - 16 kS/s per channel;
16 kS/s max aggregate for entire board |
FIFO size |
512 samples |
Input bias current |
±5nA max, ±0.5nA typical |
Input impedance |
10MΩ |
Gains |
1,2,4,8,10,20,40,80,100,200,400,800 |
Frontend bandwidth |
48kHz @ -3dB |
Common mode rejection |
100dB typical |
Oversampling ratio |
2 to 8192, selected automatically |
Accuracy |
±287.59 µV at ±10 V input range. Please see
table on page 3 for other ranges. (Typical test
results also depicte on pages two and three.) |
Isolation |
350 Vrms |
Overvoltage protection |
-40V to +55V |
Power consumption |
1.4W (stand-by); 2.2W max |
Operating temp. (tested) |
-40°C to +85°C |
Operating humidity |
95%, non-condencing |
Vibration |
IEC 60068-2-6 |
5 g, 10-500 Hz, sinusoidal |
IEC 60068-2-64 |
5 g (rms), 10-500Hz, broadband random |
Shock |
IEC 60068-2-27 |
50 g, 3 ms half sine, 18 shocks @ 6 orientations
30 g, 11 ms half sine, 18 shocks @ 6 orientations |
MTBF |
637,000 hours |
|